Bluelab pH Pen

Quick answer: to calibrate a pH pen, clean the probe, then calibrate it in fresh pH 7 solution, rinse it, and calibrate it again in fresh pH 4 solution. This is called a two point calibration. Bluelab recommends doing it once a month, and again after cleaning the probe or changing the batteries.

A pH pen drifts with use, and a pen that reads wrong is worse than no pen at all, because you end up adjusting a good feed to a bad number. Calibrating takes a few minutes and puts it right. The steps below are Bluelab's own for the Bluelab pH Pen; most pH pens follow the same pattern, but always check your meter's manual for the button presses.

What you need

Bluelab says to calibrate with fresh pH 7 and pH 4 solutions. If your readings will be higher than pH 7, you can use pH 7 and pH 10 instead.

How to calibrate a Bluelab pH Pen, step by step

  1. Clean and hydrate the probe. Bluelab says to make sure the pen is clean and hydrated before you start (see cleaning and storage below).
  2. Switch it on. Press the power/hold button.
  3. Calibrate in pH 7. Put the tip in fresh pH 7.0 solution and wait for the reading to settle between pH 6.6 and pH 7.5. Hold the Cal button until CAL shows, then let go. Zeros flash across the screen; when they stop, the screen shows pH7 and that point is done.
  4. Rinse. Rinse the tip well in clean tap water and shake off the excess.
  5. Calibrate in pH 4. Put the tip in fresh pH 4.0 solution and wait for the reading to settle between pH 3.6 and pH 4.5. Hold Cal until CAL shows, let go and wait for the zeros to finish. The screen shows pH4.
  6. Check the confirmation. A calibration mark appears at the bottom of the screen. Bluelab says it disappears after 30 days, to remind you to calibrate again.

Pour each solution into a cup rather than dipping the pen into the bottle, use it once and throw it away. Never pour used buffer back into the bottle.

How often should you calibrate a pH pen?

Bluelab recommends once a month for its pH products. It also says to calibrate:

  • before first use
  • when the calibration mark has gone from the screen (30 days after calibrating)
  • when a reading is not what you expected
  • after cleaning and hydrating the probe
  • after changing the batteries

Other makers differ, so follow your own meter's manual. For example, our Milwaukee pH51 page advises calibrating before use for the most accurate readings.

Cleaning the probe

Bluelab's method: take off the cap, fill a plastic cup with tap water and add a few drops of probe cleaner, swirl the pen in it, then rinse with fresh tap water. For heavier dirt, gently scrub the probe with a soft, clean toothbrush and rinse again. Calibrate after every clean. The Bluelab Probe Care Kit has what you need to clean, calibrate and look after a Bluelab probe.

Storing the probe so it keeps reading true

The glass bulb must never dry out. Bluelab says to keep the tip moist with KCl storage solution, adding it to the well inside the cap once a week, and never to use RO, distilled or deionised water. If the probe does dry out, Bluelab says to soak the tip in KCl storage solution for 24 hours and test it again.

Bluelab also warns not to plunge a cold probe into a hot liquid or the other way round, as sudden temperature changes can crack the glass, and not to touch the glass bulb with your fingers.

Calibrating an EC meter

EC meters are checked with a solution of known conductivity. Growth Technology EC Conductivity Standard comes in 1.413 or 2.76 mS/cm to match your meter's calibration point. Some meters need no calibration at all: Bluelab states that the Bluelab CF Truncheon does not require it, and the Bluelab Combo Meter's conductivity probe is factory calibrated.

Frequently asked questions

What can I use to calibrate a pH meter?

Fresh pH buffer solution, normally pH 7 and pH 4. Bluelab says pH 7 and pH 10 can be used instead if your readings will be above pH 7.

Do pH meters need to be calibrated?

Yes, glass probe pH meters do. Bluelab recommends monthly calibration for its pH products, and every maker gives its own interval in the manual.

Can I calibrate a pH pen with tap water?

No. Tap water is fine for rinsing between solutions, but calibration needs buffer solution of a known pH.

Why will my pH pen not calibrate?

Bluelab's steps start with a clean, hydrated probe, so clean it, soak it in KCl storage solution and try again with fresh solution. If it still will not calibrate, call us on 0121 553 7872 and we will help you check it.

What should I store my pH pen in?

pH probe storage solution (KCl). Never plain water: it draws the salts out of the probe and shortens its life.

Once your pen is reading true, our guide to reading a nutrient feed chart explains what pH and EC to aim for, and pH Up and pH Down brings your feed into range. See every pen, buffer and storage solution in our pH and EC meters range or the Bluelab range. Need a hand? Visit us at Unit 2 Black Lake, West Bromwich, call 0121 553 7872 or WhatsApp us.